Запис Детальніше

Conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination

Vernadsky National Library of Ukraine

Переглянути архів Інформація
 
 
Поле Співвідношення
 
Title Conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination
 
Creator Dzyadevich, S.V.
Zhylyak, G.A.
Soldatkin, A.P.
 
Description The characteristics of conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination have been studied. Urease has been immobilized by cross-linking with bovine serum albumin using glutaraldehyde. The resulting conductivity changes are produced by enzymatically catalyzed hydrolysis of urea. This process for both immobilized enzyme and soluble one is described by the classic laws of enzymatic kinetics. The influence of ionic strength, pH and buffer capacity of the samples on the biosensor response has been thoroughly tested. The results have been discussed concerning urea concentration analysis in human blood.
 
Date 2019-06-14T12:12:07Z
2019-06-14T12:12:07Z
1996
 
Type Article
 
Identifier Conductometric urease microbiosensor based on thin-film interdigitated electrodes for urea determination / S.V. Dzyadevich, G.A. Zhylyak, A.P. Soldatkin // Биополимеры и клетка. — 1996. — Т. 12, № 1. — С. 53-57. — Бібліогр.: 9 назв. — англ.
0233-7657
http://dspace.nbuv.gov.ua/handle/123456789/153663
577.15+573.6
 
Language en
 
Relation Биополимеры и клетка
 
Publisher Інститут молекулярної біології і генетики НАН України